| Commissioning results |
|
To illustrate the performance of the new diffractometer, we have measured a LaB6 standard with the following parameters: Beam size: 0.5 X 0.5 mm wavelength: 1.06Å Sample size: 300µm capillary Detector distance: 148mm data for 2θ values from 0° to 120° in 7 frames. Each frame is a 10° phi scan with progressive exposure times from 10 seconds to 28 seconds The data have been integrated using the software CrysAlis, and the images have subsequently been merged using photoshop.